Recall Z-0220-2020 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

LivaNova USA Inc.

SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The Stckert Heater-Cooler System is an independent (i.e. independent of the water supply) 3-circuit-heating/cooling system. The three separate water circuits have been developed to serve as: " Two circuits for the patient supplied by one tank (for the heating/cooling blanket and the heat-exchanger associated with the ox

DateNov 1, 2018
StatusTerminated
ReasonThe firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.
Root causeDevice Design
ActionOn 10/18/2018, the firm sent a certified "MEDICAL DEVICE CORRECTION" Letter to its customers informing them that due to a chemical reaction, the disinfectant/water preservative (hydrogen peroxide) used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. The letter advises customers to: -Daily Hydrogen Peroxide Monitoring: Users should monitor the hydrogen peroxide concentration in the water solution on a daily basis to verify that sufficient concentration of hydrogen peroxide is present in the water circuit of the device. A decrease in hydrogen
Quantity8671
DistributionWorldwide distribution - US Nationwide distribution including sate of PR and countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052601STOCKERT HEATER-COOLER SYSTEM 3T
Sorin Group Deutschland GmbH, Jun 6, 2006