Recall Z-0247-2020 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

TherMax Blood Warmer Unit

DateOct 30, 2019 (initiated Sep 19, 2019)
StatusTerminated
ReasonTherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
Root causeDevice Design
ActionBaxter Healthcare notified customers on about 09/19/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that specific TherMax Blood Warmers listed below may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. The affected units may not have an adequate ground path, which could cause the units to become damaged if used with improperly grounded outlets. The damage could cause the TherMax Blood Warmer units to be nonfunctional during clinical use. Instructions inclu
Quantity504 units total
DistributionNationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand
FDA recordOpen on fda.gov