Recall Z-0062-2020 ยท MOS Coil, Magnetic Resonance, Specialty

GE Healthcare, LLC

AIR Technology Anterior Array, Catalogue Numbers: (a) 5746694, (b) 5746695, (c) 5746696, a receive-only RF coil designed for use with GE 3.0T MRI systems.

DateMar 19, 2019
StatusTerminated
ReasonThe outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since the internal porous material cannot be disinfected.
Root causeDevice Design
ActionThe firm hand delivered the notices beginning on 03/19/2019. The notices described the separation of the outer seam and requested that the consignee check each coil component for tears or damage, and to discontinue use of product if product is suspected of not being in good condition. The affected products will be replaced. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative
Quantity22 units
DistributionWorldwide Distribution - US Nationwide in the states of AZ, CA, MS, NY, WI, DE, and countries of France, Germany, Japan, Netherlands, Spain, Sweden,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172695AIR Anterior Array, AIR Posterior Array
Ge Healthcare Coils (Usa Instruments, Inc.), Nov 22, 2017