Recall Z-0175-2020 ยท DSI Detector And Alarm, Arrhythmia

GE Healthcare, LLC

ApexPro Telemetry Server System. Also identified as Modification To: ApexPro Telemetry System and ApexPro FH Telemetry System - Product Usage: These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered.

DateOct 23, 2019 (initiated Aug 30, 2019)
StatusOpen, Classified
ReasonMay not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.
Root causeSoftware design
ActionGE Healthcare notified customers on about 08/30/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were made aware of the below six (6) potential safety issues along with corresponding safety instructions. Safety Issue #1: Telemetry patients monitored on a Central Station or in combination monitoring mode with a CARESCAPE B850/650/450, DASH 3000/4000/5000, or SOLAR 8000M/I bedside monitor under certain conditions can experience the following: ECG arrhythmias configured to MEDIUM (WARNING) or lower alarm priority occurring before and after a NO TELEM condition may not re-activate
Quantity22,438 (17,971 US; 4,467 OUS) devices total
DistributionWorldwide distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. The countries of ALGERIA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000779APEXPRO TELEMETRY SYSTEM
General Electric Medical Systems Information Techn, May 18, 2000
K021325MODIFICATION OF APEXPRO TELEMETRY SYSTEM
General Electric Medical Systems Information Techn, May 7, 2002
K032369MODIFICATION TO: APEXPRO TELEMETRY SYSTEM
Ge Medical Systems Information Technologies, Aug 15, 2003
K033365APEXPRO FH TELEMETRY SYSTEM
Ge Medical Systems Information Technologies, Nov 6, 2003
K080251APEXPRO TELEMETRY SYSTEMS
Ge Medical Systems Information Technologies, Jun 20, 2008