Recall Z-0173-2020 ยท GKZ Counter, Differential Cell

Abbott Laboratories

Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.

DateMay 9, 2019
StatusTerminated
ReasonThe diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.
Root causeNonconforming Material/Component
ActionThe recalling firm issued letters on 5/9/2019 to the consignees who had the installed instruments with the affected serial numbers. The letters were issued via FedEx priority overnight express.
Quantity22 analyzers with tubing
DistributionDistribution was made to GA, LA, MA, NC, NY, SC, and TX. There was no military/government distribution. Foreign distribution was made to Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081495CELL-DYN EMERALD SYSTEM
Abbott Laboratories, Feb 4, 2009