Recall Z-0527-2020 ยท MBF Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented

Zimmer Biomet, Inc.

Copeland HA Resurfacing Head, MB/HA EAS Head; Item Nos. 11-114641 11-114642 11-114643 11-114644 11-114644S 11-114645 11-114646 11-114647 11-114648 11-114661 11-114662 11-114663 11-114664 11-114664S 11-114665 11-114666 11-114667 11-114668 11-114632 11-114633 Product Usage: Partial Shoulder Replacement.

DateAug 21, 2019
StatusTerminated
ReasonElevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Root causeEnvironmental control
ActionOn September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday throug
Quantity219988 units (total)
DistributionUS Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010635INTERLOK / HA COPELAND RESURFACING HEADS
Biomet, Inc., Aug 20, 2001
K010827COPELAND MB/HA RESURFACING HUMERAL HEADS
Biomet Orthopedics, Inc., Sep 14, 2001
K051843COPELAND EAS HUMERAL RESURFACING HEADS
Biomet, Inc., Sep 29, 2005