Recall Z-0120-2020 ยท HFX Clamp, Circumcision

Centurion Medical Products Corporation

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR

DateAug 28, 2019
StatusTerminated
ReasonA mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Root causeProcess control
ActionCenturion Medical Products issued Notification via Certified Mail, Return Receipt on 8/28/19 to customers who received affected product. Letter states reason for recall, helath risk and action to take: identify inventory within your possession and cease use immediately. Complete the enclosed accountability record, whether or not you have product, and fax or email at your earliest convenience to the attention of Lisa Carpenter as directed. Upon receipt of your response Centurion Representative will contact you regarding retrieval of the product from your facility, if required. Forward this
Quantity1530 each
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K890897CENTURION CIRCLAMP