Recall Z-2756-2019 ยท LWK Razor, Surgical

Medline Industries Inc

MEDLINE ELECTRIC SURGICAL CLIPPER AND BASE, REF DYND70800 Product Usage: Medline Surgical Clipper is intended for the removal of body hair from the patient as required prior to surgery or other minor procedures.

DateSep 6, 2019
StatusOpen, Classified
ReasonThe recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base malfunction.
Root causeProcess control
ActionMedline Industries initiated the recall on 09/06/2019 by letter. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Immediately check your stock for the affected item number and lot number listed on the destruction form. Quarantine all affected product. Destroy all product belonging to the affected lot numbers listed on the destruction form. 2. Please return the completed enclosed destruction form listing the quantity of affected product destroyed. Even if you do not have any affected product please complete and return the
Quantity500 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.