Recall Z-0492-2020 ยท JQP Calculator/Data Processing Module, For Clinical Use

Beckman Coulter Inc.

AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255

DateNov 9, 2018
StatusOpen, Classified
ReasonBeckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .
Root causeOther
ActionA "Urgent Medical device Recall" letter was sent to affected consignees on November 9, 2018 via USPS. The letter informed consignees of the following: 1. If you are using the MTP kit option of the AutoMate 2500 Family systems, please ensure that your system meets the required software prerequisites stated in the Issue section. " If the requirements are met, you can safely continue using your system. " If the requirements are not met; Permanently discontinue using the MTP kit option or Contact your Beckman Coulter representative to upgrade your system to meet the required software prerequis
Quantity83 automate systems
DistributionUS: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.