Recall Z-0119-2020 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase

Horiba Instruments Incorporated

Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944

DateOct 30, 2018
StatusTerminated
ReasonPotential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).
Root causeDevice Design
ActionA "field Safety Notice" was mailed to all consignees on XX/XX/XXXX via XXXX. The field safety notice informed customer of the following: 1. If your lipase result is > 38 U/L (or the reference (normal) value of your laboratory) rerun lipase alone: - If result is still > 38 U/L(or the reference value of your laboratory), report the result as abnormal. - If result is < 38 U/L(or the reference value of your laboratory), report the result as a normal. 2. To avoid rerun, another option is to perform lipase test in batch mode (all lipase tests executed consecutively in a same run). 3. Those recom
Quantity1539 Chemistry Analyzers