Recall Z-0697-2020 ยท PEV Dialysis Administration Kit

NxStage Medical, Inc.

Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices: 1) NxStage One, Model NX1000-1; 2) NxStage VersiHD Cycler, Model Numbers: NX1000-16, and NX1000-16-A; 3) NxStage One, Model Numbers: NX1000-3 and NX1000-3-A; 4) NxStage One, Model Numbers: NX1000-10 and NX1000-10-A;

DateAug 27, 2019
StatusTerminated
ReasonThere is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.
Root causeUnder Investigation by firm
ActionOn August 27, 2019, NxStage issued URGENT MEDICAL DEVICE CORRECTION notices to customers. Actions to be Taken by Customers: For patients using NxStage System One Cycler, model NX1000-1: - At any dialysate flow rate, the system may have a fluid removal error defined as the greater of 3% of Total Dialysate Volume or 300 ml/12hr. The Total Allowable Error is generally dependent on total dialysate volume processed for this System model. For patients using less than 10L of dialysate per treatment, this model has a minimum Total Allowable Error volume of 300 ml. For patients using NxStage Sys
Quantity1803 units
DistributionAL, AZ, CA, NC & VA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050525NXSTAGE SYSTEM ONE
Nxstage Medical, Inc., Jun 24, 2005