Recall Z-2549-2019 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips North America LLC

Philips HeartStart Infant/Child SMART Pads Cartridge, Model # M5072A, UDI:00884838023758

DateAug 8, 2018
StatusTerminated
ReasonThe recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child pad cartridge should not be used on an adult because it delivers lower energy (50 J vs 150 J) than the Adult pads cartridge and may therefore not be effective on an adult.
Root causeLabeling False and Misleading
ActionA labeling specification for the HS1 is available against which labeling can be checked prior to design transfer and/or localization. This activity has been moved under regulatory affairs for review and maintenance.
Quantity181,894 Infant/Child SMART Pads Pads
DistributionUS: AK, AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, WY, Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Denmark, Dominican Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020715PHILIPS M5066A AND M5068A
Philips Medical Systems, Nov 8, 2002