Recall Z-2754-2019 · NHX Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)
Siemens Healthcare Diagnostics, Inc
Atellica IM anti-CCP IgG (aCCP) Assay Siemens Material Number: 10732998 (100 Test) Test Version1.0 - Product Usage: The ADVIA Centaur Anti-CCP IgG (aCCP) assay is for in vitro diagnostic use in the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma (K2-EDTA, lithium heparin) using the ADVIA Centaur XP system.
| Date | Aug 14, 2019 |
| Status | Terminated |
| Reason | Atellica IM anti-CCP IgG (aCCP) Test Definition (TDef) Mitigation Failure-reagent carryover mitigation for the Testosterone II (TSTII) assay is unable to be completed as defined in the aCCP Test Definition (TDef). The issue occurs on an Atellica IM Analyzer and is isolated to TDef Version 1.0. FOR OUS: Atellica IM anti-CCP IgG (aCCP) and Pregnancy-Associated Plasma Protein-A (PAPP-A) Test Definition (TDef) Mitigation Failure |
| Root cause | Under Investigation by firm |
| Action | Siemens Healthcare Urgent Medical Device Correction, AIMC 19-02.A.US sent to US customers on August 14, 2019 for delivery on August 15, 2019. Urgent Field Safety Notice, AIMC 19-02.A.OUS was sent out to the country (OUS) organizations on August 13, 2019 to customers outside the United States (US). The customer distribution included all customers that have received both Atellica IM anti-CCP IgG (aCCP) and Atellica IM Testosterone II (US and OUS) and/or both Atellica IM Pregnancy-Associated Plasma Protein-A (PAPP-A) and Atellica IM Testosterone II (OUS only) The letters inform customers that S |
| Quantity | 169 kits |
| Distribution | Worldwide distribution - US Nationwide distribution including the states of CA, FL, KY, NH, TX and countries of Austria¿¿¿¿¿¿¿¿ Curacao,¿St¿Eus Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿ |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K153551 | ADVIA Centaur Anti-CCP IgG (aCCP) assay including Reagents and Calibrators, Quality Controls and Master Curve Materials. Axis-Shield Diagnostics Limited, Aug 4, 2016 |