Recall Z-2449-2019 ยท KPR System, X-Ray, Stationary

GE Healthcare, LLC

Proteus XR/a (SlOK : K993090)

DateSep 20, 2019 (initiated Aug 9, 2019)
StatusOpen, Classified
ReasonIntermittently not receiving an audible exposure indication upon completion of an exposure on the Proteus stationary radiographic system.
Root causeDevice Design
ActionGE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to you or through a GE Healthcare field engineer site visit. A GE Healthcare representative will contact you to arrange for the correction . If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1- 800-437-1171or your local Service Representative.
Quantity647 US; 167 OUS
DistributionWorldwide - US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993090SILHOUETTE FC
Ge Medical Systems, Inc., Oct 1, 1999