Recall Z-0770-2020 ยท OBO Tomography, Optical Coherence

Carl Zeiss Meditec, Inc.

The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures incl

DateOct 16, 2018
StatusTerminated
ReasonUnder certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
Root causeDevice Design
ActionRecalling Firm issued Urgent Medical Device Recall notices to all customers dated October 16, 2018. Customers were advised to take the following actions: Until the mandatory hardware update has been performed, follow the procedure below to continue the use of the PLEX Elite 9000: When adjusting the height of the Power Table, touch only the black plastic portion of the Up/Down switch. Avoid any contact with the entire metal enclosure (including mounting screws) surrounding the black plastic Up/Down switch. Failure to follow this procedure may result in potential injury. CAUTION: Operator(s)
Quantity151 devices
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161194PLEX Elite 9000 SS-OCT
Carl Zeiss Meditec, Inc., Oct 26, 2016