ClearanceBrief
Recall Z-0289-2020 ยท DRS Transducer, Blood-Pressure, Extravascular

Argon Medical Devices, Inc

Argon Medical Devices Double Male LL Adapter, REF 040184800A, bulk, non-sterile. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.

DateOct 16, 2018
StatusTerminated
ReasonThe seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Root causeEquipment maintenance
ActionThe recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.
QuantityTotal number in recall for all products in dist: 49,130 units
DistributionDistribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K810637COBE DISPOSABLE PRESSURE TRANSDUCER
Cobe Laboratories, Inc., Apr 23, 1981