Recall Z-2761-2019 ยท GFO Activated Partial Thromboplastin

Helena Laboratories, Corp.

Cascade Abrazo aPTT Test Card, Model 5722, For Export Only. The firm name on the label is Helena Laboratories, Beaumont, TX.

DateSep 27, 2019 (initiated Dec 7, 2017)
StatusTerminated
ReasonSome of the cards show occasional shorter clot times than the expected variation resulting in a potential false negative result on the screening assay.
Root causeUnknown/Undetermined by firm
ActionThe recalling firm issued letters dated 11/12/2017 via email on 12/7/2017 explaining the issue and that they were doing a withdrawal of the lot.
Quantity21 kits
DistributionDistribution was made to Germany and Australia. There was no U.S. distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.