Recall Z-2544-2019 ยท HBG Drills, Burrs, Trephines & Accessories (Manual)

Ad-Tech Medical Instrument Corporation

AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N

DateAug 27, 2019
StatusTerminated
ReasonThe inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Root causeUnder Investigation by firm
ActionThe firm initiated the recall by letter on 08/27/2019. The firm requested the units be returned.
Quantity315 units
DistributionNY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170442Cranial Drill Bits and Accessories
Ad-Tech Medical Instrument Corporation, Sep 7, 2017