Recall Z-1165-2020 ยท HFM Monitor, Uterine Contraction, External (For Use In Clinic)
Pacific Medical Group Inc.
Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341
| Date | Sep 11, 2019 |
| Status | Open, Classified |
| Reason | Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress. |
| Root cause | Device Design |
| Action | On 09/11/19, gent Medical Device Recall notices were emailed to customers. Customers were advised of the following possible failures associated with the post-market events: 1) Fluid damage/intrusion, solder cracks, component failure, 2) Damaged membrane or cracked case, 3) Internal short from usage, broken connector from misuse, 4) Broken leads, loose coil, 5) Loose crystal from failure of adhesive, internal short or improper grounding, 6) Miscalibration of the device/waveform, 7) Overheat of device. Malfunction of the device is likely to be recognized easily because the device will not funct |
| Quantity | 466 |
| Distribution | Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K012718 | MODIFICATION TO COROMETRICS MODEL 120IS MATERNAL/FETAL MONITORING SYSTEM. Ge Medical Systems Information Technologies, Nov 13, 2001 |