ClearanceBrief
Recall Z-2556-2019 ยท DYB Introducer, Catheter

Arrow International Inc

Arrow 8.5 Fr. X 16 cm Quad-Lumen Central Venous Catheterization Kit Product Code:AK-42854-P1A - Product Usage: The Arrow Central Venous Catheters are intended to permit venous access to the central circulation by way of the femoral, jugular or subclavian veins.

DateAug 6, 2019
StatusTerminated
ReasonProducts may contain the incorrect banner card within the kit
Root causePackaging process control
ActionArrow International issued notification dated 8/6/19 stating problem, health risk and action to take: If you have affected stock in inventory, immediately discontinue use and quarantine any products with the product codes and lot numbers. 2. Inspect affected products within your control to confirm the product code on the lidstock matches the product code on the banner card of the same kit. Products confirmed to have an incorrect banner card should be returned to Arrow International. Products which properly match should not be returned. 3. If you have product with an incorrect banner card, plea
Quantity231 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall