Recall Z-2555-2019 ยท DYB Introducer, Catheter

Arrow International Inc

8.5 Fr. X 10 cm ARROW+gard Blue PSI Kit Cath-Gard for use with 7.5 - 8 Fr.Catheters Product Code: CDC-29803-1A - Product Usage: The MACTM Two-Lumen Central Venous Access Device with ARROWg+ard Blue permits venous access and catheter introduction to the central circulation. It may be inserted into the jugular, subclavian, or femoral veins. The ARROWg+ard technology is intended to help provide prote

DateAug 6, 2019
StatusTerminated
ReasonProducts may contain the incorrect banner card within the kit
Root causePackaging process control
ActionArrow International issued notification dated 8/6/19 stating problem, health risk and action to take: If you have affected stock in inventory, immediately discontinue use and quarantine any products with the product codes and lot numbers. 2. Inspect affected products within your control to confirm the product code on the lidstock matches the product code on the banner card of the same kit. Products confirmed to have an incorrect banner card should be returned to Arrow International. Products which properly match should not be returned. 3. If you have product with an incorrect banner card, plea
Quantity520 eaches
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011761MAC TWO-LUMEN CENTRAL VENOUS KIT WITH ARROWG ARD BLUE ANTIMICROBIAL SURFACE AND INTEGRAL HEMOSTASIS VALVE/SIDE PORT
Arrow Intl., Inc., Oct 31, 2001