ClearanceBrief
Recall Z-0139-2020 ยท FRN Pump, Infusion

ICU Medical Inc

Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set, 15 Micron Filter in Sight Chamber, CLAVE Secondary Port, 0.2 Micron Filter, Polyethylene Lined Light Resistant Tubing, Distal Microbore Tubing, Secure Lock 269 cm / 14 mL. 1 unit per pouch; 50 pouches per case. Not distributed in the U.S.

DateJul 29, 2019
StatusTerminated
ReasonPotential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
Root causeUnder Investigation by firm
ActionICU Medical notified customers, via Stericycle, Inc., on about 07/29/2019 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Customers were instructed to check inventory and quarantine any sets that are affected by the recall, return them to ICU Medical, inform potential users within your organization of the recall, and complete and return the provided response form. Please notify customers if affected product was further distributed and request that they contact Stericycle at 1-855-311-5437 (M-F, 8am-5pm ET) to receive a reply form. Customers were also requested to consider the following si
Quantity521 units
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. International distribution to Canada, Chile, France, Germa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.