Recall Z-2437-2019 ยท OBO Tomography, Optical Coherence

Heidelberg Engineering GmbH

SPECTRALIS with HEYEX2 image management system. a non-contact ophthalmic diagnostic imaging device.

DateJul 3, 2019
StatusTerminated
ReasonError in the default configuration which could lead to the incorrect display of patient master data.
Root causeUnder Investigation by firm
ActionFirm sent field safety notice 7/3/2019. Additionally the customer was asked to let a service technician access their software system to configure it in a way that the software error does not lead to further problems. 1. Do not drag-and-drop patient data onto the automatically created worklist partner as described above. 2. Do not execute a query or retrieve data using the automatically created worklist partner. 3. Always check whether displayed patient master data match the selected patient. In the event that displayed patient master data do not belong to the selected patient, do not cre
Quantity29
DistributionUS nationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181594Spectralis HRA+OCT and variants with OCT Angiography Module
Heidelberg Engineering GmbH, Sep 13, 2018