Recall Z-2753-2019 ยท NDC Calculator, Drug Dose

Roche Diabetes Care, Inc.

Accu-Chek Connect Diabetes Management App

DateMay 3, 2019
StatusTerminated
ReasonUsers with Android OS 8.0 and above may be unable to transfer values obtained on their meter to the App.
Root causeSoftware in the Use Environment
ActionOn 5/2/19, the firm notified affiliates of the issue via a Quality Notice. On 5/16/19, the firm issued an updated Quality Notice stating that a new version of the software, 2.1.7.2 would soon become available which would resolve the data transfer issues. Call centers were also provided with a workaround for customers who called in reporting an issue with transferring results. On 5/24/19, version 2.1.7.2 of the Connect App was released for Android on the Google Play store. This version of the app corrects the data transfer issue. At this time, users were notified via the "What's New" section o
DistributionNationwide and worldwide usage of the Connect App.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150910ACCU-CHEK Connect Diabetes Management App
Roche Diabetes Care, Inc., Jun 3, 2015