Recall Z-2367-2019 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis Q floor 10848280; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen biplane 10848355; Artis zee ceiling 10094137; Artis Q ceiling 10848281; Artis Q zeego

DateJul 30, 2019
StatusTerminated
ReasonArtis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts
Root causeProcess control
ActionSiemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice (AX011/19/S) on 8/1/19 a . The letter identifies the reason for recall, health risk and action to take: Siemens Service organization will inspect the table mainframe for possible cracks and will contact you shortly to arrange a date to perform this corrective action. Contact the service organization for an earlier appointment at 1-800-888-7436.
QuantityUS 17
DistributionUS Nationwide Distribution - AL, AR, KS, MI, MN, NC, ND, NY, OH, SC, TX, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018