Recall Z-2542-2019 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific Corporation

Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Generator

DateAug 21, 2019
StatusOpen, Classified
ReasonPotential need for device replacement - Elective Replacement Indicator (ERI) or End of Life (EOL) - earlier than expected due to compromised performance of an electrical component causing accelerated battery depletion.
Root causeComponent design/selection
ActionBoston Scientific began notifying physicians of the recall on about 08/21/2019, via "URGENT: MEDICAL DEVICE ADVISORY" letter. The physicians associated with the affected devices were provide a letter with a list of affected model/serial numbers for patients which they have implanted and/or followed or sent to their center. The letter informed physicians of the potential for accelerated battery depletion on select active worldwide EMBLEM Subcutaneous Implantable Cardioverter Defibrillators (S-ICDs). Physicians were instructed to follow up with affected units implanted in patients by enroll a
Quantity28 devices
DistributionNationwide distribution to Alabama, Arizona, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Or
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.