Recall Z-0148-2020 ยท JGS Electrode, Ion Specific, Sodium

Siemens Healthcare Diagnostics, Inc.

Siemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000 Product Usage: automated, clinical chemistry analyzer to perform in vitro diagnostic tests on clinical specimens

DateJul 30, 2019
StatusTerminated
ReasonAtellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below
Root causeSoftware design
ActionSiemens issued an Urgent medical Device Correction, (UMDC ASI19-02.A.US) via Fed Ex and Urgent Field Safety Notification (UFSN ASI19-02.A.OUS) to all affected Customers beginning on 2019-07-30. The UFSN/UMDC explain the behaviors that can occur and instructs customers on actions to be taken for the issues described in the communication. Customers are requested to return the attached effectiveness check form via fax or email within 30 days. Field Service personnel will be sent a Service Knowledge Base describing the issue and instructing them how to deal with customer questions.
Quantity1125 units
DistributionWorldwide Distribution - US Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Brazil Bulgaria Canada Chile Colombia Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Hungary India Indonesia Iran Ireland Israel Italy Japan Kuwait Latvia Malaysia Mexico Netherlands Ni
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151767Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Br
Siemens Healthcare Diagnostics, Inc., Dec 11, 2015