Recall Z-2481-2019 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
OMNIlife science Inc.
OMNIlife science, Inc. ApeX acetabular liner hip prosthesis component ApeX-LNK crosslinked Acetabular Insert, 10 deg hood; Product Code : H5-54840
| Date | Jul 23, 2019 |
| Status | Terminated |
| Reason | Acetabular liners failed the oxidation testing performed prior to release. They may experience a higher wear rate, leading to revision surgery. |
| Root cause | Nonconforming Material/Component |
| Action | OMNIlife science, Inc. sent an Urgent Medical Device Recall letter, dated July 23rd, 2019. 1. Identify the affected product from inventory and hold (quarantine) the product. 2. Return product to Omni at 480 paramount drive, Raynham, MA 02767 3. Acknowledge this notice by retuning the acknowledgement and receipt form by fax to 508-819-3390 or email to Omni-cs@coringroup.com For further information contact Christina Rovaldi, Regulatory Affairs Manager at 774-226-1847 or 800-448-6664. Return the attached Acknowledgement form by FAX to 508-819-3390 or by email to Omnics@coringroup.com. Comp |
| Quantity | 6 |
| Distribution | US nationwide distribution to states of: IL, NY, OK, MD, AL, PA, CO, FL, and UT. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K100555 | APEX-LNK ACETABULAR INSERT, E, APEX-LNK ACETABULAR INSERT, F, APEX-LNK Omni Life Science, Inc., Mar 29, 2010 |