Recall Z-2477-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Roche Diagnostics Operations, Inc.
Cobas 8000 Modular Analyzer Series: cobas 8000 ISE module part Numbers : 05964075001 cobas c 701 module part Numbers : 05641489001 cobas c 702 module part Numbers : 06473245001
| Date | Jul 18, 2019 |
| Status | Terminated |
| Reason | Quality issue with high pressure solenoid valves |
| Root cause | Device Design |
| Action | 1. For the cobas 8000 ISE, cobas c 701, and cobas c 702 modules and the cobas c 513 analyzer, stop using these units immediately. Your FES will be in contact with you to replace these affected high pressure solenoid valves immediately. 2. In the event that the high pressure solenoid valves cannot be replaced immediately, for the cobas e 601 module, follow the wash station checks outlined in this Urgent Medical Device Correction (UMDC). Perform these checks once per shift. If these checks show favorable outcomes and quality control results are within range, patient results may be reported. If |
| Quantity | 19 units |
| Distribution | State NY MT HI IN MD TX OH KS KY CA FL OR LA MO AK |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K100853 | COBAS 8000 MODULAR SERIES ANALYZER Roche Diagnostics Corp., Sep 9, 2010 |