Recall Z-2441-2019 ยท IZL System, X-Ray, Mobile

Fujifilm Medical Systems U.S.A., Inc.

Fujifilm FDR Go Plus mobile X-ray system

DateJul 19, 2019
StatusTerminated
ReasonThe graphics driver of the FDR Go PLUS might cause the appearance of a Blue Screen of Death (BSoD)
Root causeSoftware design
ActionFuji Film issued notifications via FedEx standard overnight mail on July 19, 2019 advising of the problem, health risk and action to take: During this field action, you can continue to use your current FDR Go PLUS by following these additional instructions:1. Set the monitor sleep mode to Off. Complete the Field Action Verification Form, and upon receipt FUJIFILM service personnel will contact you to schedule a visit to correct your device.FUJIFILM will correct your affected device(s) by applying a registry file to your current software version, at no cost to you.. Question contact: Field acti
Quantity174 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080701SHIMADZU MOBILE X-RAY SYSTEM MOBILEDART EVOLUTION
Shimadzu Corp., May 14, 2008