Recall Z-2499-2019 ยท HWA Impactor

Exactech, Inc.

Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use.

DateSep 10, 2019 (initiated Jul 24, 2019)
StatusTerminated
ReasonThe threaded mechanism of the Threaded Offset Cup Impactor may be gritty and difficult to turn potentially due to entrapped blast media.
Root causeNonconforming Material/Component
ActionExactech notified customers on about 07/24/2019 via "URGENT MEDICAL DEVICE RECALL" letter. Instructions included to immediately cease distribution or use of the affected devices, provide the information to customers if further distributed, identify and quarantine any affected devices, and complete and return the provided Recall Acknowledgement Form and Recall Inventory Response Form to Exactech via email at recalls@exac.com. On 09/09/2019, Exactech expanded the recall to include an additional lot. Customers were notified of the expansion on about 09/09/2019.
Quantity22 devices
DistributionUS distribution to AR, CA, FL, IL, IN, NY, SC, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.