Recall Z-2551-2019 ยท NDW Assay, Porphyrin, Spectrophotometry, Lithium

Sentinel CH SpA

MULTIGENT Lithium-intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 8L25-30

DateJun 28, 2019
StatusTerminated
ReasonAdded an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results
Root causeUnder Investigation by firm
ActionSentinel issued Urgent Field Safety Notice letter on 6/28/19 via email to the distributor. The Distributor Abbott notified customers by letter Urgent Field Safety Notice Product Correction dated July 11, 2019 advising of the problem, health risk and action to take: If the LDH reagent is installed on your instrument, please follow the instructions 1. Update the Lithium (LN 8L25-30) assay parameters to add the new SmartWash. For additional information, see the ARCHITECT System Operations Manual, Section 2, Configure the SmartWash settings (c System. Install the Alinit y c Lithium assay file ver
Quantity15919 kits
DistributionNationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY JORDAN KUWAIT LEBANON LUXEMBOU
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070987SENTINEL LITHIUM ASSAY
SENTINEL CH. SpA, Aug 2, 2007