ClearanceBrief
Recall Z-2438-2019 ยท JTC Device, Microtiter Diluting/Dispensing

Becton Dickinson & Co.

BD Kiestra InoqulA Software Version 20.3.IVD automated specimen processing Catalog Number: 447202

DateJul 26, 2019
StatusTerminated
ReasonAn anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
Root causeSoftware design
ActionBD initiated a field correction via letter on July 26, 2019 and BD Technical Service has completed the correction (software upgrade) for all six (6) affected US customers. Letter identifies issue, health risk and customer action to take. BD is in the process of evaluating the log files of customers using the InoqulATM / InoqulA+TM software version 20.3 to determine if any users potentially experienced mismatch between a specimen and plate. BD will contact all customers using InoqulATM / InoqulA+TM software version 20.3 to provide the outcome of this evaluation. Contact: BD Customer/Technical S
Quantity4 units
DistributionIllinois, Iowa, Missouri, New York, Ohio, Tennessee World Wide (5): Singapore, Taiwan, France, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.