Recall Z-0160-2020 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Siemens Medical Solutions USA, Inc

Mammomat Revelation with software version VC10 and Biopsy Option Intended Use: For mammography exams, screening, diagnostics, biopsies and dual energy procedures

DateJul 23, 2019
StatusTerminated
ReasonAn error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases
Root causeSoftware design
ActionSiemens issued the customer notification on 7/23/19 via XP018/19/S. The Advisory Notice provides the problem, health risk and action to take: provided steps to the operator on how to easily release patient and proceed with exam. Siemens is developing a corrective action to resolve the issue, which will be released via Update Instruction XP021/19/S.
Quantity139 WW (33 US)
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173408Mammomat Revelation
Siemens Medi Cal Solutions, Inc., Mar 21, 2018