Recall Z-1906-2020 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci SP Instrument Arm Drape. REF 430013-12 QTY 1 Pouch

DateOct 25, 2018
StatusTerminated
ReasonThe instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier.
Root causeDevice Design
ActionOn 10/25/2018, the firm sent an "Urgent Medical Device Correction" Notification Letter to Customers via FedEx (to OUS customers - emailed and hand-delivered) are asked to take actions during draping to avoid snagging the drape during installation and to continue following the instructions in the System User Manual. Representatives will be available by phone to answer questions related to the device correction. Acknowledgement form included with the recall letter is requested to be completed and returned to ISI asking that appropriate personnel be informed about the device correction.
Quantity1818 pouches
DistributionUpdated information provided on 03/18/2020: FL, OH, TN, NJ, PA, MD, NY, CA, MN, NV, OR, HI, VA, TX, AL, CA, IL, CN, and NC. OUS: APAC (Asia Pacific) - South Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173906da Vinci SP Surgical System, EndoWrist SP Instruments, and Accessories
Intuitive Surgical, Inc., May 31, 2018