Recall Z-0024-2020 ยท ETN Stimulator, Nerve

Stryker Instruments Div. of Stryker Corporation

C2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System is intended for intra-operative monitoring and stimulation for localization and identification of cranial and peripheral motor and mixed motor sensory nerves during surgery.

DateJul 31, 2019
StatusTerminated
ReasonStryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device, which may result in use errors potentially causing or contributing to nerve injury.
Root causeDevice Design
ActionOn July 31, 2019, the firm distributed an Urgent Medical Device Recall Notification to its consignees to inform them of the recall. Customers were asked to do the following: 1. Immediately review the Recall Notification and this Business Reply Form (BRF). 2. Inform all users of the Stryker C2 NerveMonitor of the information contained in this recall notification. 3. Check all treatment areas, operating rooms and/or storage areas for affected products. 4. Determine a plan to transition to an alternative commercially available product. Document your transition date on the BRF. This transition d
Quantity135
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152505C2 NerveMonitor System
Inomed Medizintechnik GmbH, Feb 1, 2016