ClearanceBrief
Recall Z-2347-2019 ยท LCO Platelet Factor 4 Radioimmunoassay

Aniara Diagnostica LLC

ZYMUTEST HIA MonoStrip, IgGAM ELISA kit, Model No. RK041D, for in vitro diagnostic use.

DateJul 19, 2019
StatusTerminated
ReasonResidual crystallization of the microplate and recurrent negative controls out of range.
Root causeUnder Investigation by firm
ActionThe firm notified customers of the recall via Urgent Medical Device Correction emails dated July 26, 2019. Customers were informed of the recall and asked to destroy the affected kits in stock and indicate the quantity in the Awareness Acknowledgement Form, which was to be signed and returned to the local distributor. If you have any questions regarding this recall, please contact Aniara at (513) 360-6377, or CMurphy@aniara.com.
Quantity46
DistributionDistribution to US states of NC, NH, NY, and OH, and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071255ZYMUTEST HIA IGGAM, ZYMUTEST IGG, MODEL RK040D, RK040A
Hyphen Biomed, Sep 7, 2007