ClearanceBrief
Recall Z-2287-2019 ยท OZN C. Difficile Toxin Gene Amplification Assay

Luminex Corporation

ARIES C. difficile Assay , REF 50-10018, UDI # 00840487100059

DateAug 15, 2019 (initiated Feb 20, 2019)
StatusOpen, Classified
ReasonThrough internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Root causeProcess control
ActionOn February 20, 2019, the firm sent a "Urgent: Medical Device Recall/Field Safety Notice" via email to all consignees. The customer notification informed customers of the following: 1. As an immediate action, Luminex is requesting that customers discontinue use of and discard the products listedin the recall notice immediately or by the Discontinue Use By date. Contact Luminex Global Support Services to obtain replacement material. Material that is still within its Discontinue Use By date may be used at the customers discretion. 2. Complete the Acknowledgment and Receipt form. You must
DistributionUS: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171441ARIES C. difficile Assay Complete Kit, ARIES C. difficile Assay Protocol File Kit, ARIES C. difficile Assay Kit (24 cass
Luminex Corporation, Jul 21, 2017