Recall Z-0419-2020 ยท DWJ System, Thermal Regulating

Cincinnati Sub-Zero Products LLC, a Gentherm Company

CoolBlue Hyper-Hypothermia System- body hypothermia system Model: Innercool - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.

DateJul 2, 2019
StatusTerminated
ReasonWarnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
Root causeLabeling False and Misleading
ActionGentherm issued URGENT Medical Device Correction, Blanketrol II letters were mailed by USPS starting July 25th and Blanketrol III letters were mailed by USPS starting July 30th. Letters stated problem, health risk and action to take: INSTRUCTIONS TO CUSTOMERS: 1) Access updated manuals and ensure that obsolete manuals are removed from service. Updated manuals may be accessed via www.gentherm.com or physical copies may be requested from Gentherm Medical, LLC at (513)326-5320. 2) Ensure that all users are informed of the contents of this letter. If you have further distributed this product, plea
Quantity5291
DistributionWorldwide distribution - US Nationwide and countries of Afghanistan, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bhutan, Brazil, Brunei, Canada Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, France, Germany, Guatemala, Hong Kong, India, Indonesia, Israel,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101589BLANKETROL III, MODEL 233 AND COOLBLUE, MODEL 25-01 HYPER-HYPOTHERMIA SYSTEMS
Cincinnati Sub-Zero Products, Inc., Sep 20, 2010