ClearanceBrief
Recall Z-0012-2020 ยท KJJ Cleanser, Root Canal

Inter-Med Llc

Vista Dental Products Sodium Hypochlorite - 6%, 10 - 12cc Pre-filled Syringes, REF 502300

DateOct 1, 2019 (initiated Jun 27, 2019)
StatusTerminated
ReasonThe affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
Root causeProcess control
ActionThe firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.
Quantity1992 units
DistributionUS, South Africa, Switzerland, Poland, Italy, Singapore, Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082470SODIUM HYPOCHLORITE 3 & 6% SOLUTION AND 6% WITH WETTING AGENTS TO LOWER SURFACE TENSION MARKETED AS CHLOR-XTRA
Inter-Med, Inc., Apr 3, 2009