Recall Z-2275-2019 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

NeuroLogica Corporation

HS70A Diagnostic Ultrasound System Version 2.01.00, 2.01.01, 2.01.02, 2.01.03, 2.01.04, 2.01.05, 2.01.06, 1.00.06, 1.00.07, 1.00.08, 1.00.09, 1.00.10, 1.00.11, 1.00.12

DateJul 25, 2019
StatusTerminated
ReasonThere is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.
Root causeSoftware design
ActionOn July 25, 2019, Neurological Corporation a subsidiary of Samsung Electronics issued Urgent Field Safety Notices to customers. Customers were advised to the following actions: 1) If the user wants to decrease the Dopper SV (Sample Volume) size set by himself/herself in the Doppler Only Mode, the user should press "set" button to update the PW Doppler mode. 2) A copy of the Field Safety Notice should be passed on to all who need to be aware within your organization. 3) Contact your local Samsung representative if you have questions. 4) Return the completed Customer Reply Form to To
Quantity3582 worldwide
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173713HS70A Diagnostic Ultrasound System
Samsung Medison Co., Ltd., Mar 8, 2018