Recall Z-0117-2020 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

Aestiva 7100, Model Numbers: 1. 1006-9320-000 2. 1006-9321-000 3. 1006-9320-000-001101 4. 1006-9021-000 5. 1006-9012-000 6. 1006-9111-000 7. 1006-9321-000-007639 8. 1006-9321-000-000023 9. 1006-9321-000-015180 10. 1006-9320-000-010369 11. 1006-9321-000-019486 12. 1006-9321-000-007601 13. 1006-9321-000-009266 14. 1006-9321-000-008526 15. 1006-9321-000-116507 16. 1006-9321-000-009244 17. 1006-9321

DateJul 19, 2019
StatusTerminated
ReasonCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
Root causeSoftware in the Use Environment
ActionThe firm sent an urgent medical device correction notice to their consignees by letter on 07/19/2019. The letter explained the issue and requested that the consignee ensure that the devices are connected to a sufficiently secure terminal server.
Quantity473 devices
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000706DATEX-OHMEDA AESTIVA/5 WITH 7100 VENTILATOR
Datex-Ohmeda, Inc., May 11, 2000