Recall Z-0115-2020 ยท CBK Ventilator, Continuous, Facility Use

GE Healthcare, LLC

Aestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n) 1006-9120-000 o) 1006-9027-000 p) 1006-9003-000 q) 1006-9025-000

DateJul 19, 2019
StatusTerminated
ReasonCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
Root causeSoftware in the Use Environment
ActionThe firm sent an urgent medical device correction notice to their consignees by letter on 07/19/2019. The letter explained the issue and requested that the consignee ensure that the devices are connected to a sufficiently secure terminal server.
Quantity7168 devices
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023366DATEX-OHMEDA 7900 VENTILATOR ENHANCEMENTS
Datex-Ohmeda, Inc., Oct 7, 2003