Recall Z-0114-2020 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

Aestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077 f) 1006-9310-000-015243 g) 1006-9310-000-017602 h) 1006-9310-000-103785 I) 1006-9310-000-025109 j) 1006-9310-000-009650 k) 1006-9310-000-015224 l) 1006-9310-000-031881 m) 1006-9310-000-031854 n) 1006-9310-000-026571

DateJul 19, 2019
StatusTerminated
ReasonCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
Root causeSoftware in the Use Environment
ActionThe firm sent an urgent medical device correction notice to their consignees by letter on 07/19/2019. The letter explained the issue and requested that the consignee ensure that the devices are connected to a sufficiently secure terminal server.
Quantity502 devices
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050055GE DATEX-OHMEDA AESTIVA/5 MRI ANESTHESIA SYSTEM
Datex-Ohmeda, Inc., Feb 4, 2005