Recall Z-2362-2019 ยท NHX Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)

Roche Diagnostics Operations, Inc.

Roche Elecsys Anti-CCP Immunoassay Catalog # 05031656160

DateAug 23, 2019 (initiated Jul 16, 2019)
StatusTerminated
ReasonPerformance issues with certain lots of the Elecsys Anti-CCP assay on the cobas e 411 analyzer; MODULAR ANALYTICS E 170 module; and cobas e 601, 602, and 801 modules with plasma samples.
Root causeUnder Investigation by firm
ActionCustomers will be instructed to only use serum samples when testing with one of the affected lots numbers. Re-measure the sample if an implausible result occurs while using plasma samples, and a high Anti-CCP result does not match the patients clinical picture. To follow the work-around provided in the UMDC concerning pre-analytical sample handling for plasma samples. Perform maintenance according to the operator manual (e.g., Liquid Flow Cleaning [LFC]) to ensure proper functioning of the analyzer or module. Customers will be reminded that sample quality can be affected by fibrin clots an
Quantity550
DistributionAZ CA FL HI IA IL IN KY LA MA MI MO MT NJ NY OK OR PA PR SC TX VA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081338ROCHE ELECSYS ANTI-CCP IMMUNOASSAY, ROCHE ELECSYS PRECICONTROL ANTI-CCP
Roche Diagnostics, Sep 11, 2008