Recall Z-2345-2019 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

MAGNETOM Prisma magnetic resonance diagnostic devices Model #10849582

DateJul 9, 2019
StatusTerminated
ReasonPotential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
Root causeDevice Design
Action1. review the System Owner Manual chapter 1.3.3 Quench emergency plan and the section regarding defective vent line. 2. follow the standard emergency procedures you have in place in the event of system failures. 3.A Siemens Service Technician will conduct an inspection to verify that your system has a vertical outlet. If a vertical outlet is present, no further action is required. In the event your system is equipped with a horizontal outlet, the Siemens Service Technician will ensure that the port for the drop-off plate is adequately blanked off or sealed. 4.service organization will contact
Quantity17
DistributionAL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132119MAGNETOM PRISMA, MAGNETOM PRISMA FIT
Siemens Medical Solutions USA, Inc., Nov 22, 2013