ClearanceBrief
Recall Z-2319-2019 ยท JJH Clinical Sample Concentrator

Qiagen Sciences LLC

QIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids-IVD) Model Number: 9001297 Product Usage: The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids. It is intended to be used only in combination with QIAsymphony kits indicated for use with the

DateJul 2, 2019
StatusOpen, Classified
ReasonQIAsymphony software version 5.0.3 - Software issue incorreclty assigns the 2D bar code eluate ID to the sample within the results file. If 2D bar code eluate IDs are not checked against the sample ID prior to downstream application the incorrect tube may be selected and it could have the potential to lead to delayed results or incorrect patient reporting
Root causeSoftware design
ActionQiagen notified Customer on July 2, 2019 via email and provided reason for recall, heath risk and action to take: Discontinue use of the workflow in the following specific condition: QIAsymphony software version 5.0.3 when using the 2D bar code integration feature with continuous loading. Please note that the 2D bar code labware, continuous loading, and Software 5.0.3 can still be used independently of the integration feature. Complete and sign the included Acknowledgement of Receipt form and email it to QIAGEN Technical Services at techservice-eu@qiagen.com or fax it to +49 (0)2103-29-2240
Quantity417
DistributionWorldwide Distribution - US Nationwide Foreign: Algeria Australia Austria Azerbaijan Belgium Brazil Canada China Czech Republic Denmark Finland France Germany Greece India Israel Italy Japan Jordan Korea, Republic of Kuwait Latvia Mexico Netherlands Norway Pakistan Poland Republic of Korea Russian F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.