Recall Z-0046-2020 ยท NOT Microscope, Automated, Image Analysis, Operator Intervention

Leica Biosystems Imaging, Inc.

Aperio CS2 CE IVD scanner instrument, scanner part #23CS100CE.

DateOct 4, 2019 (initiated Dec 10, 2018)
StatusTerminated
ReasonThe incorrect sensor was used in the assembly of the camera in the scanner instrument which can potentially generate whole slide images that impact analysis algorithm results, as well as potentially impact the visual appearance of whole slide images.
Root causeComponent design/selection
ActionThe consignees were notified beginning on 12/10/2018 via letter delivered by email.
Quantity3 devices
DistributionThere was no U.S. distribution. Foreign distribution was made to South Korea and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141109APERIO EPATHOLOGY EIHC IVD SYSTEM
Leica Biosystems Imaging, Inc., Jul 29, 2014