Recall Z-2278-2019 ยท HSB Rod, Fixation, Intramedullary And Accessories

Pega Medical Inc.

SLIM Handle - catalogue #: SLM-HND100 a component of Simple Locking Intramedullary (SLIM) System

DateMay 29, 2019
StatusTerminated
ReasonHandle could jam with the Driver due to the detachment of an internal Connector Ring.
Root causeComponent design/selection
ActionAn e-mail was sent on May 25, 2019 to all their consignees with the following instructions. Identify the device with the same catalogue number and lot number and return the device to Pega Medical Inc. as soon as possible to the address shown below using the FedEx Account number provided. Contact Enrique Garcia at (450) 688-5144 ext. 242 or egarcia@pegamedical.com for more information
Quantity19
DistributionCA, DE, FL, IL, NC, NE, NY and TN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143355The Simple Locking IntraMedullary (SLIM) System
Pega Medical, Inc., Jul 10, 2015