Recall Z-1997-2019 ยท OWB Interventional Fluoroscopic X-Ray System

Philips Medical Systems Nederlands

Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2

DateNov 30, 2018
StatusTerminated
ReasonPhilips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.
Root causeDevice Design
ActionOn November 30, 2018, Philips initiated a software update in the field. Among other performance improvements , this software update included the correction for poor image quality under very specific situations.
Quantity229
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017